Tip: Box pressure of 8 Pa or less is recommended because higher pressures will decrease the exhaust fan flow rate.
Box pressures below 3 Pa are less accurate because small fluctuations in pressure will affect the flow a lot. One Pa is a
practical limit for measuring flow, below which readings cannot be taken. Try the same fan with different holes to see
the different results.