
Page 43 of 66
©Retrotec Inc. 2012
6 Field Calibration
Gauge Calibration
To verify the calibration of a gauge, the easiest method is to compare the readings of one channel with
respect to the other channel. If the pressures are equal, then it is likely the gauge is accurate, because
the chance of both channels being out of calibration by the same amount is very small.
To perform a cross port check
1. Set both channels to measure pressure in Pascals.
2. Turn Auto-Zero off
3. Connect the Input port of Channel A to the Input port of Channel B using a small piece of
tubing. The measured pressure on both channels should be within 1% of each other.
4. Disconnect the tubing, and then connect the Reference ports of both channels using the
same piece of tubing. The measured pressure on both channels should be within 1% of each
other.
The same procedure can be followed to compare two gauges. The reading on both gauges
should be the same.
Figure 8. Gauge field check configurations (cross-channel checks (top), cross-gauge checks (bottom)).
Comentarios a estos manuales