Retrotec DM-2A Manual de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 66
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
Page 50 of 66
©Retrotec Inc. 2012
Water can also cause significant errors if a drop is pulled into the pressure tubes. It is very easy to get a
water drop inside the end of a tube. Simply drag it over a shallow puddle of water, and a drop of water
will get pulled into the tube by capillary action. Notice in the Figure 13 how one drop of water in the
end of the tube is creating 190 Pascal pressure. It is possible to complete a Door Fan test with a water
drop in the end of the tube. It will appear as if the fan is creating a pressure in the building, but in fact
the results are way off. Particularly unusual behavior in the gauge could mean that there is water in the
end of the tube.
Be careful that no one steps on the tubes, or creates a kink in any of the lines while performing a
test. If this happens, wait a few seconds for the effects to subside before taking any readings from
the DM-2.
Whip the end of the tube around to remove excess water. If this doesn’t work (in extreme cases),
water can be blown out from the other end. Be sure to not have the gauge connected when
blowing water from the tube. Water can cause either a positive or a negative pressure, based on
which way the drop of water is pulling. If a drop of water is facing downward, it pulls out of the
tube, creating a negative pressure (as displayed in Figure 13). Point the tube upward, and the water
drop falls back into the tube, creating a positive pressure.
Figure 13. Water in the reference tube (left); high fixed pressure due to water (right).
7.7 Wrong Flow Range or Device selected
If the results just don't make sense, check to make sure the correct Flow Range and device are
selected. Each Device and Flow Range configuration has its own individual flow curve that is used to
convert fan pressure into flow, and into the results selected under the Mode menu. If a test is
completed with the incorrect Device and/or Flow Range selected, it is possible to salvage the data by
using the customized flow charts provided in Appendix A.
To determine correct flow if wrong Flow Range/device was selected:
1. Select the fan pressure for which a corrected flow value is needed.
2. Find this fan pressure value in the left column of the flow conversion tables in Appendix A.
3. Move across the row and record the flow value beneath the column heading with the actual
Flow Range that was used.
Note: Retrotec 2000 and 3000 series fans have the same flow curves.
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 65 66

Comentarios a estos manuales

Sin comentarios