7.1 DM-2 Rechargeable Battery Problems
The DM-2 is designed to recharge NiMH batteries in the gauge, eliminating the need to purchase a
stand-alone battery charger. The charging cycle is designed to fully discharge batteries before
recharging them. This can cause problems if the DM-2 is connected to a power plug for only brief
periods of time.
If the gauge exhibits any of the following conditions, it may be a problem with the rechargeable
batteries:
Short battery life
Will not turn on
Keys will not respond
Loss of power during test
The solution is to reset gauge and charge the batteries, or switch to non-
rechargeable batteries.
To charge the batteries:
1. Connect the AC power supply if rechargeable NiMH batteries are already installed.
2. Press the reset button on the back of the DM-2. The gauge turns on.
Note: DM-2's that have firmware prior to version 2.28 will not have a reset
button. To reset the gauge, remove all power connections, and the batteries,
and wait at least 5 minutes. Then re-install the batteries and reconnect the
charger.
3. Press [Exit] to show the main screen.
4. Display indicates batteries are charging.
5. Charge for at least 8 hours, or until battery icon stops moving.
Problems can be avoided by taking a few simple steps.
Leave DM-2 disconnected from a power supply until the batteries are
depleted to one quarter power or less.
Do not short-charge batteries by interrupting a charge in progress.
Replace batteries with new AA NiMH batteries after 2 years.